Adverse Event Report: 1224623

Robotic surgeries involving the da Vinci Surgical System have been known to cause many adverse events, including fatality. If you believe your injuries, or a loved one's death, was caused by complications due to the use of this medical device, you could be entitle to compensation.

Our defective medical devices lawyers are here to help provide the answers and legal guidance you need. Contact us today for a 100 percent free, no-obligation case evaluation.

INTUITIVE SURGICAL, INC. DA VINCI ENDOWRIST CADIERE FORCEPS FORCEPS, DISSECTING, ROBOTIC SURGICAL    
Model Number 420049 VER-05
Event Date 09/10/2008
Event Type  Malfunction  
Event Description

During surgical procedure a splinter size piece of graphite material was noted in the patient's abdomen. It was apparent that the piece broke free from the robotic wristed instrument. This happened during the removal and insertion of the instrument from the robot arm. Doctor was aware and indicated that it would not have any adverse effect on the patient. Intuitive surgical has since provided the hospital with an inspection procedure for our 8mm cannulae, which involves a visual inspection and the use of a "gage pin", also provided by intuitive surgical. The procedure is recommended to be performed prior to sterilization and prior to surgery to detect damaged cannulae.

 

Gordon & Doner - For The Injured®

1 (855) 722-2552

Gordon & Doner, P.A.

4114 Northlake Blvd
Palm Beach Gardens, FL 33410
Phone: 561-333-3333

Gordon & Doner - Davie

10650 Florida 84 #210
Davie, FL 33324
Phone: 754-333-3333

Gordon & Doner - Stuart

729 SW Federal Highway #212
Stuart, FL 34994
Phone: 772-333-3333

Privacy Policy | Sitemap | © 2010 - 2017 Gordon & Doner, P.A. - PMP Marketing Group